Specifications


Appearance
Color: Matt black

Ergonomic design: The height adjustment range is 125mm, and the screen can be switched between the horizontal screen and the vertical screen, rotated 45 degrees left and right, and tilted 5 degrees forward and 15 degrees backward

VESA installation: pitch-row 100 x 100 (mm)

Buttons: Mechanical buttons (MODE, MENU, UP, DOWN, BACK, and Power)

 

Screen
Backlight: LED

Screen size: 21.3-inch

Resolution: 2560 x 2048

Gray scale/color display: 10 bits

Pixel pitch: 0.165 x 0.165 mm

Brightness: 3000 cd/m2 (typ)

Contrast ratio: 2000:1

Response time: 25 ms

Visual angle: 178 , 178

Image
DICOM PART14: Yes

Data processing and display: (14-bit LUT data processing, 10-bit gray scale and color display)

Image mode: DICOM300, DICOM350, DICOM400, DICOM450, DICOM500, DICOM550, DICOM600, DICOM650, DICOM700, DICOM custom, user mode, and ambient light mode

Gamma curve: DICOM/CT/MRI/DSA/DR/CR/ MM/1.8/2.0/2.2/2.4/2.6

Color temperature (non-adjustable): Specified (default)

Brightness adjustable: 0-100 stepping adjustment

Fast and stable backlight function: Support (backlight sensor)

Long-time stable backlight function: Support (backlight sensor)

Ambient light compensation: Support (ambient light sensor)

Saving energy control for body sensing: Support (human sensor)

Auto menu rotation: Support (gravity sensor)

One-click lamp box function: Support

Maximum power: < 80 W

 

Power supply specification
Power indicator: Green and orange

Standby power: < 0.5 W

Power requirement (BOM including the adapter): Medical adapter, AC 100-240 V 50/60Hz input, 24 V output

Input and output control
Input: DVI-D x 1, DP x 1

Output: DP daisy chain

USB: One upstream interface and two downstream interfaces

Series interface: Upstream USB interface reused

DDC communication: Support

Fixing equipment: Base

 

Display improvement
DICOM conformance error rate: Within 5% (allowing 3% fluctuation)

 

Others
Trademark on the front shell: Hisense Medical

Boot screen: Hisense Medical image