HMD2G21S

21.3” Diagnosis Monitor


Hisense clinical review displays are strictly conformed to DICOM standard, which can be widely used in clinical diagnosis, video browsing and image audit purposes.

21S1
21S1
21S2
21S2

Specifications


Appearance
Color: Dark gray matt

Ergonomic design: The height adjustment range is 125mm, and the screen can be switched between the horizontal screen and the vertical screen, rotated 45 degrees left and right, and tilted 5 degrees forward and 15 degrees backward.

VESA installation: pitch-row 100 x 100 (mm)

Buttons: "Touch button (LIGHT, MODE, UP (highlighted), DOWN, OK, BACK, and POWER)"

 

Screen
Backlight: LED

Screen size: 21.3-inch

Resolution: 1600 x 1200

Gray scale/color display: 8 bits

Pixel pitch: 0.27 x 0.27 mm

Brightness: 1900 cd/m2 (typ)

Contrast ratio: 1800:1

Response time: 19 ms

Visual angle: 178 , 178

Image
DICOM PART14: Yes

Data processing and display: (14-bit LUT data processing, 8-bit gray scale and color display)

Image mode: User mode (Custom), DICOM200, DICOM250, DICOM300, DICOM350, DICOM400, DICOM450, DICOM500, DICOM600, DICOM700, and DICOM custom

Gamma curve: DICOM, CT, MRI, DSA, Gamma1.8, Gamma2.0, Gamma2.2, Gamma2.4, and Gamma2.6

Color temperature (non-adjustable): Specified (default)

Brightness adjustable: 0-100 stepping adjustment

Fast and stable backlight function: Support (backlight sensor)

Long-time stable backlight function: Support (backlight sensor)

Ambient light compensation: Support (ambient light sensor)

Saving energy control for body sensing: Yes

Auto menu rotation: Support (gravity sensor)

One-click lamp box function: Support

Maximum power: < 80 W

 

Power supply specification
Power indicator: Green and orange

Standby power: < 0.5 W

Power requirement (BOM including the adapter): Medical adapter, AC 100-240 V 50/60Hz input, 24 V output

Input and output control
Input: DVI-D x 1, DP x 1

Output: DP daisy chain

USB: One upstream interface and two downstream interfaces

Series interface: Upstream USB interface reused

DDC communication: Support

Fixing equipment: Base

 

Display improvement
DICOM conformance error rate: Within 5% (allowing 3% fluctuation)

 

Others
Trademark on the front shell: Hisense Medical

Boot screen: Hisense Medical image

Configuration GPU: (It is uniformly produced and delivered in the packaging box, and added to the accessory details of the user manual and quick installation manual.)

 

LCD control Software
Local brightening function: Implemented by shortcut keys and the mouse

Quality control software: (Standalone edition)

Display control: (Brightness adjustment)

Health remind: Yes

Key Features


DICOM Part14 Standard
DICOM Part14 Standard
The DICOM-compliant display accurately represents the subtle shading of medical images, and guarantees consistant display over time in different terminals.
Wide Viewing Angle
Wide Viewing Angle
The wide viewing angle of 178 degrees can ensure the users from different angles can obtain the same image quality without colour cast.
Eye-Protection and Energy-Saving
Eye-Protection and Energy-Saving
Continuously monitor ambient brightness and adjust screen accordingly in time to reduce visual fatigue and enhance reading experiences.
Intelligent Gamma Calibration
Intelligent Gamma Calibration
Hisense patented intelligent GAMMA calibration technology is suitable for reading grayscale and colour images on the same screen simultaneously, which will improve reading accuracy.
High Grayscale and Colour Processing
High Grayscale and Colour Processing
Using high-gray processing technology, you can obtain more smoother and precise images.
Rapid Brightness Stability Technology
Rapid Brightness Stability Technology
Quickly reach the brightness steady state within 10s after starting up, to ensure that the doctor can quickly carry out the diagnosis work.